Troponin Rapid Test Accuracy
INTENDED USE
ONE STEP TROPONIN I TEST IS AN IMMUNOCHROMATOGRAPHY BASED ONE STEP IN VITRO TEST. IT IS DESIGNED FOR QUALITATIVE DETERMINATION OF CARDIAC TROPONIN I (CTNL) IN HUMAN WHOLE BLOOD, SERUM OR PLASMA AS AN AID IN THE DIAGNOSIS OF MYOCARDIAL INFARCTION.
MATERIALS PROVIDED
1. Test cards individually foil pouched with a desiccant
2. Plastic dropper
3. Package insert
MATERIALS REQUIRED BUT NOT PROVIDED
1. Timer/clock
2. Pipette
3. Controls
ASSAY PROCEDURE
1. Read package insert carefully before testing. Allow the test devices, whole blood, serum or plasma to equilibrate to room temperature (15-30°C) prior to testing. Do not open pouches until ready to perform the assay.
2. Remove the test device from the foil pouch and use it as soon as possible.
3. Place the test device on a clean and level surface. Hold the dropper provided vertically and transfer 3 drops of specimen (100µl) to the specimen well (S) in the test device.
4. Wait for the red line(s) to appear. The result should be read between 10 to 15 minutes.
Accuracy
Cut off | Special |
0.5ng/ml | 99.9% |
0.2ng/ml | 97% |
0.1ng/ml | 93% |
Interference testing
The following substances were added to the negative control and 0.5ng/ml Troponin I spiked serum samples. No interference was found with any of the substances at the following concentrations:
substance | concentration |
Bilirubin | 10 mg/ml |
Cholesterol | 800 mg/ml |
Hemoglobin | 250 mg/ml |
Triglyceride | 250 mg/ml |
sTnI | 1000ng/ml |
cTnT | 1000ng/ml |
cTnC | 1000ng/ml |
Troponin Rapid Test Accuracy